Hello Du,
Do you only encounter this arcing issue when running the microscope in
environmental SEM mode? If so, it probably means your chamber pressure is
too high and the water vapor (or whatever gas you're flowing) is ionizing.
Try reducing the pressure in your chamber.
Good luck,
Brian C. Stahl
Graduate Student Researcher
UCSB Materials Research Laboratory
[email protected] / [email protected]
Cell: (805) 748-5839
Office: MRL 3117A
On Wed, Mar 11, 2009 at 3:30 PM, Qinghong Du wrote:
> Hi,
>
> We have a Philips XL30 ESEM system. Recently, the system only can work
> under 10kV. Over 10kV, there is arching issue in the HTSU unit. Does anybody
> have this experience and any idea to fix it?
>
> Thanks a lot.
>
> Du